日本財団 図書館


5.2.15 Electrostatic discharge immunity test (IEC60098-504/ 5)
 
5.2.15.1 Test procedure
 
The equipment is at an operating condition and apply the electrostatic discharge immunity test as specified in Table 8. The response point of the specimen shall be measured in accordance with 5.1.3 after conditioning.
 
Table 8 - Parameters for electrostatic discharge immunity test
Contact discharge 6kV
Air discharge 8kV
Interval between single discharges 1 sec.
No. of Pulses 10 per polarity
 
NOTE For detailed test methods, refer to Level 3 of IEC 61000-4-2.
 
5.2.15.2 Requirements
 
Performance Criterion B (see 5.2.21). The ratio of the response points Dmax/Dmin shall not be greater than 1,26.
 
5.2.16 Radiated radio frequency immunity test (IEC60098-504/ 5)
 
5.2.16.1 Test procedure
 
The equipment is at an operating condition and apply the [electrostatic discharge] immunity test as specified in Table 9. The response point of the specimen shall be measured in accordance with 5.1.3 after conditioning.
 
Table 9 - Parameters for radiated radio frequency immunity test
Frequency range 80 MHz - 2 GHz
Modulation 80% AM at 1 kHz
Field strength 10 V/m
Frequency sweep rate ≤1.5 x 10-3 decades/sec. (or 1%/3sec.)
 
5.2.16.2 Requirements
 
Performance Criterion A (see 5.2.21). The ratio of the response points Dmax/Dmin shall not be greater than 1,26.
 
5.2.17 Conducted low frequency immunity test (IEC60098-504/5)
 
5.2.17.1 Test procedure
 
The equipment is at an operating condition and apply the [electrostatic discharge] immunity test is carried out as specified in Table 10. The response point of the specimen shall be measured in accordance with 5.1.3 after conditioning (values in round brackets are shown where the rated frequency of the equipment is 50 Hz).
 
Table 10 - Parameters for conducted low frequency immunity test
Frequency range 60 Hz-12 kHz (50 Hz-10 kHz)
Test voltage AC 10% of supply voltage 60 Hz - 900 Hz (50 Hz - 750 Hz)
10 to 1□ of supply voltage 900 Hz - 6 kHz (750 Hz - 5 kHz)
10% of supply voltage 6 kHz - 12 kHz (5 kHz- 10 kHz)
DC 10% of supply voltage (at least 3 V) 50 Hz - 10 kHz
Maximum power 2W
 
5.2.17.2 Requirements
 
Performance Criterion A (see 5.2.21). The ratio of the response points Dmax/Dmin shall not be greater than 1,26.
 
5.2.18 Conducted high frequency immunity test (IEC60098-504/ 5)
 
5.2.18.1 Test procedure
 
The equipment is at an operating condition and apply the (electrostatic discharge] immunity test is carried out as specified in Table 11. The response point of the specimen shall be measured in accordance with 5.1.3 after conditioning.
 
Table 11
 
Parameters for conducted high frequency immunity test
Frequency range 10 kHz - 80 MHz
Modulation 80% AM at 1 kHz
Amplitude 3 V rms
 
If for tests of equipment an input signal with a modulation frequency of 1 kHz is necessary a modulation frequency of 400 Hz should be chosen.
 
NOTE For detailed test methods, refer to Level 2 of IEC 61000-4-6.
 
5.2.18.2 Requirements
 
Performance Criterion A (see 5.2.21). The ratio of the response points Dmax/Dmin shall not be greater than 1,26.
 
5.2.19 Burst/Fast transient immunity test (IEC60098-504/ 5)
 
5.2.19.1 Test procedure
 
The equipment is at an operating condition and apply the [electrostatic discharge] immunity test as specified in Table 12. The response point of the specimen shall be measured in accordance with 5.1.3 after conditioning.
 
Table 12 - Parameters for burst/fast transient immunity test
Single pulse time 5 nS (between 10% and 90% value)
Single pulse width 50 nS (50% value)
Amplitude line on power supply port/earth: 2 kV
line/line on I/O data control and signal lines: 1 kV
Pulse period 300 mS
Burst duration 15 mS
Duration 5 min./polarity
 
NOTE For detailed test method refer to Level 3 of IEC 61000-4-4.
 
5.2.19.2 Requirements
 
Performance Criterion A (see 5.2.21). The ratio of the response points Dmax/Dmin shall not be greater than 1,26.
 
5.2.20 Surge immunity test (IEC60098-504/5)
 
5.2.20.1 Test procedure
 
The equipment is at an operating condition and apply the [electrostatic discharge] immunity test is carried out as specified in Table 13. Where power and signal lines are identical, the test procedure is to be in accordance with Figure 3. The response point of the specimen shall be measured in accordance with 5.1.3 after conditioning.
 
Table 13 - Parameters for surge immunity test
Pulse rise time 1.2μS (between 10% and 90% value)
Single pulse width 50μS (50% value)
Amplitude line/earth: 1 kV
line/line: 0.5 kV
Reputation rate at least 1 pulse/min.
No. of pulses 5 per polarity
 
NOTE For detailed test methods refer to Level 2 of IEC 61000-4-5.
 
5.2.20.2 Requirements
 
Performance Criterion A (see 5.2.21). The ratio of the response points Dmax/Dmin shall not be greater than 1,26.
 
Figure 3 - Example of surge immunity test circuit
 
5.2.21 Performance criteria for [EMC] test
 
5.2.21.1 Performance criterion B
 
The equipment under test continues to operate as intended after the test. During the test, degradation or loss of function or performance which is self recoverable is permissible, but no change of actual operating state or stored data is permissible.
 
5.2.21.2 Performance criterion A
 
The equipment under test continues to operate as intended during and after the test. No degradation of performance or loss of function shall be permitted as defined in the technical specification published by the manufacturer.
 
 
6.1 Each detector shall be clearly marked with, or supplied with, the following information:
 
a) the number of this standard;
 
b) the name or trademark of the manufacturer or supplier;
 
c) the model designation (type or number);
 
d) the classification of the detector, e.g. Class 1;
 
e) some mark(s) or code(s), (e.g. a serial number or batch code) by which the manufacturer can identify, at least, the date or batch and place of manufacture and the version number(s) of any software, contained within the detector;
 
f) the wiring terminal designations;
 
g) the angle of reception as determined in 5.2.3;
 
h) the operating wavelength band(s) e.g. UV, IR.
 
6.2 For detachable detectors, the detector head shall be marked with at least a), b), c), d) and e), and the base shall be marked with at least b), c) (i.e. its own model designation) and f).
 
6.3 Where any marking on the device uses symbols or abbreviations not in common use then these shall be explained in the data supplied with the device.
 
6.4 The marking shall be visible during installation of the detector and shall be accessible during maintenance.
 
6.5 The markings shall not be placed on screws or other easily removable parts.







日本財団図書館は、日本財団が運営しています。

  • 日本財団 THE NIPPON FOUNDATION