Reference document 1
Performance test of analyzer at equipment manufacturing factory, etc.
1. Test conditions: The performance test conditions of the analyzer are as follows.
(1) Ambient temperature: Temperature between 5 and 35℃, fluctuation within +/-5℃
(2) Humidity: Relative humidity within 85%
(3) Atmospheric pressure: 95 to 106kPa (950 to 1060mbar), with fluctuation within 5%
(4) Power supply voltage: Rated voltage
(5) Power supply frequency: Rated frequency
(6) Warm-up time: Time described in operation manual
(7) Gas used for test: Standard gas,span gas, zero gas, and nitrogen monoxide. The type of gas and applicable test item are as per Table 1.
Table 1.